CoB1 - 12 um DPW Held at Ground

at 0krad Mean Threshold = 10.39, MPv Noise = 5.43

Date

Dose Rate

Time of Exposure

Total Exposure

Integrated Exposure

Run Numbers

Mean Threshold

MPV Noise

Comments

02/03/11

98 rad/s

2h50m15s

1 Mrad

1 MRad

 

14.01

5.68

Some Broken Bonds when removed from freezer

When readout using _runNumberInSequence=66 got a strange pattern and thought this was due to broken bonds. It is infact due to the runNumberInsequence chosen not the broken bonds. CoB1 was then allowed to anneal for 24, 48, and 168 hours to see the differences

Date

Annealing State

Annealing Time

RunNumbers

Mean Threshold

Mean Noise

Comments

03/03/11

Ground

24hours

912154 - 912195

13.23

5.72

 

04/03/11

Ground

48hours

912195 - 912236

13.03

5.73

 

09/03/11

Ground

168hours

912238 - 912279

12.77

5.75

 

31/03/11

Ground

3WeeksFrozen

912280 - 912321

12.86

5.78

After 3 weeks in the freezer took scan to see the effects of being frozen

31/03/11

Ground

resultofscan

912223 - 912264

12.85

5.77

Took second scan directly afterwards to see the effects of annealing whilst scanning

Following this annealing it was decided to take the sensor to higher doses and see what happened

Date

Dose Rate

Time of Exposure

Total Exposure

Integrated Exposure

Run Numbers

Mean Threshold

MPV Noise

Comments

05/04/11

98 rad/s

1h25m0s

0.5 Mrad

1.5 MRad

912365 - 912406

13.83

5.81

 

05/04/11

0 rad/s

0h0m0s

0.0 Mrad

1.5 MRad

912407 - 912448

13.89

5.79

Note:second scan taken directly aftwerwards to see the effect of annealing straight after exposure

07/04/11

98 rad/s

2h50m15s

1 Mrad

2.5 MRad

912449 - 912490

14.19

5.95

 

14/04/11

98 rad/s

2h50m15s

1 Mrad

3.5 MRad

912491 - 912532

14.63

6.09

 

21/04/11

98 rad/s

4h05m0s

1.5 Mrad

5.0 MRad

914812 - 914853

75.51

6.36

Sensor had higher dose to get to 5Mrad before easter
wrongcfg file so mean and noise different so 5 MRad not comparable to 3.5

03/05/11

5.0 MRad

915004 - 915045

13.90

5.72

5Mrad scan conducted after annealing for 168 hours with wrong cfg so cannot be trusted

03/05/11

5.0 MRad

915046 - 915087

13.22

6.02

5Mrad scan conducted after annealing for 168 hours taking in all pixels
shows beam spot well

Noticed that there was much increased threshold for 5Mrad run and found that wrong cfg file used so it cannot be compared to earlier dose data. The annealing studies were however conducted with the same cfg file so annealing study still good

Following the easter bank holiday the sensor was removed from the freezer and allowed to anneal

Date

Annealing State

Annealing Time

RunNumbers

Mean Threshold

Mean Noise

Comments

21/04/11

Ground

0hours

914812 - 914853

75.51

6.36

wrongcfg file so mean and noise different so 5 MRad not comparable to 3.5

03/05/11

Ground

24hours

914854 - 914894

74.01

6.30

 

04/05/11

Ground

48hours

914896 - 914937

73.94

6.30

 

09/05/11

Ground

168hours

914938 - 914979

73.72

6.30

 

31/05/11

Ground

672hours

915552 - 915593

73.43

6.32

 

CoB2 - 12 um DPW powered

at 0krad mean threshold = 8.85, MPV noise = 5.74

Date

Dose Rate

Time of Exposure

Total Exposure

Integrated Exposure

Run Numbers

Mean Threshold

MPV Noise

Comments

15/08/11

60 rad/s

5m33s

20 krad

20krad

921574 - 921594

7.76

5.77

 

15/08/11

60 rad/s

5m33s

20 krad

40krad

921595 - 921615

5.83

5.84

 

15/08/11

60 rad/s

5m33s

20 krad

60krad

921658 - 921578

3.87

5.88

 

15/08/11

60 rad/s

5m33s

20 krad

80krad

921679 - 921699

1.68

5.91

 

16/08/11

60 rad/s

5m33s

20 krad

100krad

921706 - 921726

0.35

5.93

 

16/08/11

60 rad/s

5m33s

20 krad

120krad

921727 - 921747

-1.61

5.93

 

16/08/11

60 rad/s

5m33s

20 krad

140krad

921790 - 921810

-3.33

5.98

 

16/08/11

60 rad/s

5m33s

20 krad

160krad

921811 - 921831

-4.73

6.00

 

17/08/11

60 rad/s

5m33s

20 krad

180krad

921832 - 921852

-5.71

6.01

 

17/08/11

60 rad/s

5m33s

20 krad

200krad

921911 - 921916

-4.97

7.10

6 scans to test
8k errors in sensorload

17/08/11

 

 

 

200krad

921917 - 921922

-6.55

6.03

6 scans after 4.5 hours 2k errors

17/08/11

 

 

 

200krad

921923 - 921943

-5.83

5.99

21 scans after 5 hours (proper scan)

18/08/11

60 rad/s

5m33s

20 krad

220krad

921992 - 921997

-4.56

6.15

6 scans to test after 690mins
errors in sensorload

18/08/11

 

 

 

220krad

922097 - 922117

-3.58

6.04

after 24 hours full scan was possible

Now as the damage is annealing away before can scan it is pointless dosing higher so now going to anneal

Date

Annealing State

Annealing Time

RunNumbers

Mean Threshold

Mean Noise

Comments

18/08/11

Powered

0hours

922097 - 922117

-3.58

6.04

 

19/08/11

Powered

12hours

922118 - 922138

-3.71

5.99

 

19/08/11

Powered

24hours

922139 - 922159

-3.50

5.98

 

20/08/11

Powered

36hours

922160 - 922180

-3.48

5.98

 

20/08/11

Powered

48hours

922181 - 922201

-3.44

5.97

 

21/08/11

Powered

60hours

922202 - 922222

-3.30

5.95

 

21/08/11

Powered

72hours

922223 - 922243

-3.09

5.97

 

24/08/11

Powered

138hours

922445 - 922465

-

 

 

30/08/11

Powered

282hours

922782 - 922802

-1.30

6.01

 

CoB3 - 12 um DPW Held at Ground

at 0krad Mean Threshold = 10.81, MPv Noise = 5.48

Date

Dose Rate

Time of Exposure

Total Exposure

Integrated Exposure

Run Numbers

Mean Threshold

MPV Noise

Comments

10/05/11

60 rad/s

4h38m20s

1 Mrad

1 MRad

915216 - 915257

13.57

5.86

No Broken bonds (smile)

11/05/11

60 rad/s

4h38m20s

1 Mrad

2 Mrad

915258 - 915299

13.71

6.11

Sensor expose 11th tested 12th kept in freezer

13/05/11

60 rad/s

4h38m20s

1 Mrad

3 Mrad

915300 - 915341

14.07

6.26

Sensor left out of freezer ~1h after exposure finished
tested 16th after being frozen over weekend

18/05/11

60 rad/s

4h38m20s

1 Mrad

4 Mrad

915510 - 915551

14.43

6.47

Sensor draw current of 0.889A but no errors from SensorLoadv1.1

19/05/11

60 rad/s

4h38m20s

1 Mrad

5 Mrad

915636 - 915677

14.65

6.48

Sensor left out for 45 mins after scan finsihed
scans not analysed until 06/06/11

After 6 weeks in the freezer due to traveling etc I know hope to conduct annealing studies on this sensor. How accurate they will be I am unsure but hopefully not much change over the 6 weeks in the freezer

Easy way to check.. perform a scan and compare to 5Mrad data before doing annealing

Date

Annealing State

Annealing Time

RunNumbers

Mean Threshold

Mean Noise

Comments

06/07/11

Ground

0hours

916944 - 916985

89.38

6.18

Due to sensor frozen for 6 weeks make a scan to know where to start annealing from
can be compared with 5Mrad data
used wrong cfg file so need to always sue this one for annealing

07/07/11

Ground

25hours

916986 - 917027

89.79

6.18

 

08/07/11

Ground

48hours

917028 - 917069

86.91

5.98

middle two sections do not seem to be working

13/07/11

Ground

168hours

917280 - 917321

89.93

6.22

middle two sections do not seem to be working

03/08/11

Ground

672hours

919344 - 919385

90.22

6.23

 

CoB4 - 12 um DPW Held at Ground

at 0krad Mean Threshold = 9.63, MPv Noise = 5.68

Date

Dose Rate

Time of Exposure

Total Exposure

Integrated Exposure

Run Numbers

Mean Threshold

MPV Noise

Comments

16/05/11

60 rad/s

0h55m30s

0.2 Mrad

0.2 MRad

915342 - 915383

11.36

5.85

Still seems to be large jump in 1st 200 krad exposure

16/05/11

60 rad/s

0h55m30s

0.2 Mrad

0.4 MRad

915384 - 915425

11.76

5.93

 

17/05/11

60 rad/s

0h55m30s

0.2 Mrad

0.6 MRad

915426 - 915467

12.10

6.01

 

17/05/11

60 rad/s

0h55m30s

0.2 Mrad

0.8 MRad

915468 - 915509

12.17

6.04

 

24/05/11

60 rad/s

0h55m30s

0.2 Mrad

1 MRad

915594 - 915635

12.05

6.07

Sensor been frozen for a week no sign of annealing

27/05/11

60 rad/s

2h19m00s

0.5 Mrad

1.5 MRad

916642 - 916683

n/a

n/a

sensor out 2 hours longer than needed whilst trained on machine
left a week in freezer before testing
ground wire left on so results are scrap

The fractional changes for CoB3 and CoB4 at 1 Mrad seem to be in agreeance. However 1500 krad exposure is scrap for many reasons. Therefore this sensor will now be abandoned

CoB6 - 12 um DPW Held at Ground

at 0krad Mean Threshold = 9.35, MPv Noise = 5.52

Date

Dose Rate

Time of Exposure

Total Exposure

Integrated Exposure

Run Numbers

Mean Threshold

MPV Noise

Comments

11/07/11

60 rad/s

0h55m30s

0.2 Mrad

0.2 MRad

917070 - 917112

9.87

5.72

 

11/07/11

60 rad/s

0h55m30s

0.2 Mrad

0.4 MRad

917112 - 917153

10.14

5.84

CoB6 returned to freezer and analysed on 12/07

12/07/11

60 rad/s

0h55m30s

0.2 Mrad

0.6 MRad

917154 - 917195

10.31

5.92

 

12/07/11

60 rad/s

0h55m30s

0.2 Mrad

0.8 MRad

917196 - 917237

10.19

5.94

Analysed 13/07 after being frozen over night

13/07/11

60 rad/s

0h55m30s

0.2 Mrad

1.0 MRad

917238 - 917279

10.38

6.00

 

13/07/11

60 rad/s

4h38m20s

1.0 Mrad

2.0 MRad

917322 - 917363

10.12

6.28

 

20/07/11

60 rad/s

4h38m20s

1.0 Mrad

3.0 MRad

917364 - 917405

11.00

6.49

 

28/07/11

60 rad/s

4h38m20s

1.0 Mrad

4.0 MRad

917406 - 917447

11.02

6.67

 

28/07/11

60 rad/s

4h38m20s

1.0 Mrad

5.0 MRad

917448 - 917489

10.93

6.78

scan conducted same day as 4MRad

following this dose CoB6 was then allowed to anneal for 168 hours

Date

Annealing State

Annealing Time

RunNumbers

Mean Threshold

Mean Noise

Comments

28/07/11

Ground

0hours

917448 - 917489

10.93

6.78

 

03/08/11

Ground

24hours

919302 - 919343

118.47

6.60

sensor removed from freezer 02/08/11
wrong cfg file again!

09/08/11

Ground

48hours

920941 - 920982

11.25

6.62

sensor removed from freezer day before

14/08/11

Ground

166hours

921532 - 921573

11.23

6.59

 

CoB7 Powered

at 0kRad Mean threshold = 10.90, MPV noise = 5.65

Date

Dose Rate

Time of Exposure

Total Exposure

Integrated Exposure

Run Numbers

Mean Threshold

MPV Noise

Comments

29/07/11

60 rad/s

0h55m30s

0.2 Mrad

0.2 MRad

917616 - 917657

-25.54

0

Huge number of cfg errors so scans are scrap

due to the huge number of errors 55Fe test conducted for 24 hours to see if pixels still readout

Date

Annealing State

Annealing Time

RunNumbers

Mean Threshold

Mean Noise

Comments

03/08/11

Powered

24hours

919386 - 919427

 

 

scan after Fe55 test

Sensor then left for 160 hours at ground before another Fe55 test for 24 hours. Then gave sensor another 100kRad and lots of cfg errors again

Date

Dose Rate

Time of Exposure

Total Exposure

Integrated Exposure

Run Numbers

Mean Threshold

MPV Noise

Comments

12/08/11

60 rad/s

0h27m45s

0.1 Mrad

0.1 MRad

 

Huge number of cfg errors so cannot scan

Keeping CoB7 powered and monitoring the number of cfg errors with time and SRAM voltage

Time Powered

Number of Errors

0hours

 

25hours

 

48hours

 

72hours

 

CoB8 Powered

at 0kRad Mean threshold = , MPV noise =

CoB9 Powered

at 0kRad mean threshold = 11.29, MPV noise = 5.70

Date

Dose Rate

Time of Exposure

Total Exposure

Integrated Exposure

Run Numbers

Mean Threshold

MPV Noise

Comments

09/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.02 MRad

920815 - 920856

9.44

5.78

 

09/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.04 MRad

920857 - 920898

6.91

5.87

 

09/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.06 MRad

920899 - 920940

7.29

5.88

laser left on so no dose given

09/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.06 MRad

920983 - 920992
920995 - 921026

4.66

5.94

 

09/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.08 MRad

921027 - 921068

2.06

6.00

 

10/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.10 MRad

921069 - 921110

-0.47

6.10

 

10/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.12 MRad

921111 - 921152

-2.73

6.13

 

10/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.14 MRad

921153 - 921194

-4.16

6.12

 

10/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.16 MRad

921195 - 921236

-5.93

6.15

1 bad row but not enough for noise pickup
post scan sensorload all problems gone

11/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.18 MRad

921238 - 921279

-5.35

6.22

8k cfg errors
down to 500 after 3 hours so took the scan

12/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.20 MRad

921322 - 921363

-4.19

6.14

30k cfg errors
down to 500 after 18 hours so took the scan

now annealing sensor powered

Date

Annealing State

Annealing Time

RunNumbers

Mean Threshold

Mean Noise

Comments

13/08/11

Powered

2hours

921364 - 921405

-4.11

6.14

 

13/08/11

Powered

5hours

921406 - 921447

-3.72

6.15

 

14/08/11

Powered

12hours

921448 - 921489

-3.33

6.14

 

14/08/11

Powered

24hours

921490 - 921531

-3.09

6.11

 

15/08/11

Powered

48hours

921616 - 921647

-2.19

6.15

 

16/08/11

Powered

72hours

921748 - 921789

-1.75

6.13

 

17/08/11

Powered

96hours

921869 - 921910

-1.22

6.17

 

18/08/11

Powered

120hours

921950 - 921991

-1.02

6.16

 

CoB12 18um hi-res powered

Sensor from ALICE people. Problem with edge effects so using 21 runs with runnner=10

at 0kRad Mean threshold = 11.89, MPV noise = 5.59

Date

Dose Rate

Time of Exposure

Total Exposure

Integrated Exposure

Run Numbers

Mean Threshold

MPV Noise

Comments

18/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.02 MRad

922013 (21)

9.44

5.70

 

18/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.04 MRad

922034 (21)

6.16

5.77

 

18/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.06 MRad

922055 (21)

3.19

5.84

 

18/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.08 MRad

922076 (21)

-0.23

5.86

 

22/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.10 MRad

922244 (21)

-1.95

5.88

 

22/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.12 MRad

922265 (21)

-4.41

5.87

 

22/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.14 MRad

922286 (21)

-6.22

5.90

 

22/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.16 MRad

922307 (21)

-7.92

5.87

 

22/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.18 MRad

922328 (21)

-9.06

5.89

scanned 30 mins after powered due to 500 errors (SRAM time)

23/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.20 MRad

922349 (21)

-9.81

5.99

500 errors, 1 column

23/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.22 MRad

922370 (6)

-11.62

6.03

1700 errors, 3 columns 60m after exposure

23/08/11

 

 

 

0.22 MRad

922397 (21)

-8.94

5.85

1 bad column 180m after exposure

24/08/11

60 rad/s

0h05m33s

0.02 Mrad

0.24 MRad

922418 (6)

-4.19

6.88

1700 errors, 3 columns 60m after exposure

24/08/11

 

 

 

0.22 MRad

922397 (21)

-6.37

5.97

1 bad column 180m after exposure

now annealing sensor powered

Date

Annealing State

Annealing Time

RunNumbers

Mean Threshold

Mean Noise

Comments

25/08/11

Powered

12hours

922466 (21)

-5.71

5.93

 

25/08/11

Powered

24hours

922487 (21)

-6.51

5.90

 

26/08/11

Powered

36hours

922508 (21)

-6.82

5.90

 

26/08/11

Powered

48hours

922529 (21)

-6.51

5.90

 

27/08/11

Powered

60hours

922550 (21)

-6.51

5.90

 

27/08/11

Powered

72hours

922571 (21)

-6.22

5.88

 

28/08/11

Powered

84hours

922592 (21)

-6.17

5.89

 

28/08/11

Powered

96hours

922613 (21)

-6.13

5.90

 

29/08/11

Powered

108hours

922634 (21)

-6.11

5.87

 

29/08/11

Powered

120hours

922655 (21)

-6.14

5.90

 

30/08/11

Powered

132hours

922676 (21)

-6.02

5.88

 

30/08/11

Powered

144hours

922761 (21)

-5.73

5.94

 

note: Mean Threshold and Mean Noise correspond to the sensor wide values and not pixel by pixel changes

  • No labels